Resolution problems



LevelBasic
It is very important to locate the problem before installing a new column. Keep in mind that use of guard columns will prolong column lifetime and avoid many problems of contamination.

Reduced resolution / separation.

 

Reduced resolution / separation
Possible cause Solution
Extra column band broadening. Make sure the tubing length and tubing id are kept to a minimum.
Check if the detector flow cell volume is smaller than 0.1 x peak volume.
Finally check the influence of the detector's time constant; keep this as low as possible (< 0.5 s.)
Improper eluent flow rate. The best eluent flow rate is found just above the optimum of the h-u curve. The use of smaller particles usually results in a shallower slope of the h-u curve. Check the system for solvent leakage and check whether the pump is delivering the eluent with a constant flow.
Improper ph or ionic strength of eluent. The ph strongly influences the concentration of the neutral acidic compounds in the eluent and therefore strongly affects retention behaviour.
Packing bed contamination.

High column pressure accompanying resolution loss.
Particulate accumulation on column inlet frit or top of the stationary phase
Sample: filter or centrifuge samples
Eluent: filter eluent, cap off eluent réservoirs
System: flush all lines and pumps, install in-line system filter

Eluent too aggressive (ph too low or too high) Adjust the mobile phase pH within the column pH limits.
 

Click here to go to the Home-page