Resolution problems
Nico Vonk, Avans+, Breda, The NetherlandsLevelBasic
It is very important to locate the problem before installing a new column. Keep in mind that use of guard columns will prolong column lifetime and avoid many problems of contamination.
Reduced resolution / separation.
| Possible cause | Solution | |
| Extra column band broadening. | Make sure the tubing length and tubing id are kept to a minimum. Check if the detector flow cell volume is smaller than 0.1 x peak volume. Finally check the influence of the detector's time constant; keep this as low as possible (< 0.5 s.) | |
| Improper column temperature stability. | In such cases the use of a column thermostat is strongly advised. | |
| Improper eluent flow rate. | The best eluent flow rate is found just above the optimum of the h-u curve. The use of smaller particles usually results in a shallower slope of the h-u curve. Check the system for solvent leakage and check whether the pump is delivering the eluent with a constant flow. | |
| Insufficient equilibration time with mobile phase. | Use longer equilibration time for the column. Check baseline and retentions stability before starting calibration and injection of samples. | |
| Improper ph or ionic strength of eluent. | The ph strongly influences the concentration of the neutral acidic compounds in the eluent and therefore strongly affects retention behaviour. | |
| Unstable gradient or mixing. | Check proportional valves or mixing chamber of the gradient pump. | |
| Packing bed contamination. |
High column pressure accompanying resolution loss. | |
| Accumulation of proteins in biological samples | Microbial growth in samples . Refresh samples or improve stockage conditions. | |
| Eluent too aggressive (ph too low or too high) | Adjust the mobile phase pH within the column pH limits. | |
| Organic contamination | Fats, oils, lipids in samples Non-specific organics from sample or improperly prepared eluents |





